Automation, convenience, speed and standardization of results

Features

  1. ESR test performed on same tubes in which blood samples taken (edta/cbc).
  2. Test can be done on small volume of blood (1.5 ml)
  3. ESR test results on a compact 7″ touch screen designed for seamless interaction and clear visualization.
Principle Infrared LED detection
Display 4.2" color touch screen
Analysis time 30 min or 60 min
Pre indication of results After 12 minutes
After 12 minutes 1-140mm/hr
Reading resolution +/-0.1mm
Accuracy +/- 1 mm
Reproducibility C. V. < 5%
Memory 9999 sample results
Samples code entry Manual entry & Barcode scanning
Printer Built-in thermal printer
Test tube Vacuum tube with 3.8% sodium citrate
Results Westergren method RT & Auto compensation to 18°C as per manley table mm/hr
Reading channels 10/20
Loading pattern Random access
Temperature 15°-30°C
Electrical specification 110-230V AC 50Hz/60Hz
Barcode scanner interface Via USB port up to 16 digits alfa numeric
LIS Via RS-232, HL7 protocol
Dimension 323mm(L) X 196mm(W) X 187mm(H)
Weight 3.8kg
Standards & Regulations ISO 13485, USFDA registered, CDSCO registered, CE certified
IEC 61010-1:2010/AMD 1:2016, IEC 61010-1:2010, IEC 61010-2-101:2018,
IEC 61326-1:2020, IEC 61326-2-6:2020

Distictive Advantage